The effect of implant size and device keel on vertebral compression properties in lumbar total disc replacement

  • Joshua D. Auerbach, MD

      Affiliations

    • Department of Orthopaedics, Bronx-Lebanon Hospital Center, Albert Einstein College of Medicine, Bronx, NY 10457, USA
    • Department of Orthopaedic Surgery, McKay Orthopaedic Research Laboratory, University of Pennsylvania, Philadelphia, PA 19104, USA
    • Corresponding Author InformationCorresponding author. Department of Orthopaedics, Bronx-Lebanon Hospital Center, Albert Einstein College of Medicine, 1650 Grand Concourse, 7th Floor, Bronx, NY 10457, USA. Tel.: (609)-202 6173.
  • ,
  • Carrie M. Ballester, MSE

      Affiliations

    • Department of Orthopaedic Surgery, McKay Orthopaedic Research Laboratory, University of Pennsylvania, Philadelphia, PA 19104, USA
  • ,
  • Frank Hammond, MSE

      Affiliations

    • Department of Orthopaedic Surgery, McKay Orthopaedic Research Laboratory, University of Pennsylvania, Philadelphia, PA 19104, USA
  • ,
  • Ehren T. Carine, BS

      Affiliations

    • Department of Orthopaedic Surgery, McKay Orthopaedic Research Laboratory, University of Pennsylvania, Philadelphia, PA 19104, USA
  • ,
  • Richard A. Balderston, MD

      Affiliations

    • Booth, Bartolozzi, and Balderston Orthopaedics, Pennsylvania Hospital, Philadelphia, PA 19104, USA
  • ,
  • Dawn M. Elliott, PhD

      Affiliations

    • Department of Orthopaedic Surgery, McKay Orthopaedic Research Laboratory, University of Pennsylvania, Philadelphia, PA 19104, USA

Received 22 July 2009; accepted 13 January 2010.

Abstract 

Background Context

Vertebral end plate support is necessary for successful lumbar total disc replacement (TDR) surgery. Failure to achieve anterior column support as a result of lumbar TDR device undersizing could lead to implant subsidence and fracture.

Purpose

The purpose of the study was to examine the compressive biomechanical behavior of the vertebral end plate with varying sizes of disc replacement implants.

Study Design

The study design comprises a biomechanical investigation using a human cadaveric lumbar spine model.

Methods

Fifty-six vertebrae with intact posterior elements were prepared from 13 fresh frozen lumbar spines. Peripheral quantitative computed tomography was performed to assess regional bone density. Vertebrae were potted and subjected to nondestructive compression testing with a small, medium, and large custom-made implants with the footplate geometry of the ProDisc-L TDR (Synthes Spine, West Chester, PA, USA) system and having no keel. Failure testing was performed using the ProDisc-L implant with an intact keel. Pressure sensor film was used to assess contact pressure and distribution.

Results

There was a linear correlation between percent coverage of the end plate and implant-end plate stiffness (p=.0001) and an inverse correlation with displacement (p=.01). The difference in implant-end plate stiffness between small-medium, medium-large, and small-large implants was 10.5% (p=.03), 10.2% (p=.02), and 19.6% (p<.0001), respectively. Failure analysis revealed similar trends for implant sizing, but only bone density was found to significantly correlate with failure properties (r=0.76, p<.0001). There was a significant reduction in implant-end plate stiffness of 18% when the keel was intact compared to without the keel (range 6–27%, p=.0008). Pressure film analysis revealed that the implant was loaded peripherally and did not have central contact during nondestructive loading. There was a trend toward greater contact pressure with the small implant when compared with the medium implant (p=.06) and the large implant (p=.06).

Conclusions

Although larger implants reduce end plate displacement, increase apparent implant-end plate stiffness, increase the implant-end plate contact area, and decrease the peak contact pressures, low bone density reduces failure properties. The keel introduces a reduction in stiffness to the implant-end plate interface, the clinical significance of which is currently unknown.

Keywords: Lumbar total disc arthroplasty, End plate, Compression, Mechanics, Keel

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 FDA device/drug status: not applicable.

 Author disclosures: JDA (consulting, Synthes Spine, DePuy Spine, Paradigm Spine, LLC, research support: staff and/or materials, Synthes Spine, Fonar, Inc.; grants, DePuy Spine); RAB (consulting, Synthes; speaking and/or teaching arrangements, Synthes; trips/travel, Synthes); DME (grants, AO Spine Foundation).

PII: S1529-9430(10)00044-6

doi:10.1016/j.spinee.2010.01.008

The Spine Journal
Volume 10, Issue 4 , Pages 333-340, April 2010