Lumbar intersegmental spacing and angulation in the modified lateral decubitus position versus variants of prone positioning
Abstract
Background Context
Interspinous process devices represent an emerging treatment for neurogenic intermittent claudication resulting from lumbar spinal stenosis. Most published descriptions of the operative technique involve treatment of patients in the modified lateral decubitus knee-chest position (modified lateral decubitus), and yet many surgeons have begun to perform the procedure in various prone positions. The patient's positioning on the operating room table seems likely to influence resting interspinous distance, and thus implant sizing and possibly the risk of intraoperative spinous process fracture. The intersegmental lumbar effect of variants on operative prone positioning compared with the modified lateral decubitus position has not been studied.
Purpose
We performed this study to determine the comparative differences in interspinous distance and intersegmental angulation effected by the lateral decubitus knee-chest position and the variants on prone positioning used in practice.
Study Design/Setting
Experimental human radiographic study.
Patient Sample
Twenty healthy male volunteers with a mean age of 43.6±10.8 years (range, 24–63), without chronic back pain, symptoms of neurogenic claudication, or history of lumbar surgery were enrolled.
Outcome Measures
Interspinous distance, anterior and posterior disc heights, disc angulation were measured on PACS monitor.
Methods
Lateral X-rays were taken of the lower lumbar spine in each of four different surgical positions (modified lateral decubitus, Andrews frame, Wilson frame, and Jackson frame). Statistical analysis was performed on the resultant data points to assess the significance of the effect of the position of the subject on intersegmental spacing and angulation.
Results
The 20 enrollees had a mean age of 43.6±10.8 years (range, 24–63). The mean interspinous distance at the L4–L5 level was greatest on the Andrews table (23.5±8.3
mm) followed by the modified lateral decubitus position (19.6±5.1
mm), the Wilson frame (15.6±4.6
mm), and then the Jackson frame (10.1±4.7
mm; significantly less than all other positions p≤.036). Mean segmental extension at the L4–L5 level was least in the modified lateral decubitus position (−0.1°±2.9°); this was statistically similar to extension on the Andrews table (1.5°±4.7°, p=1.0), but significantly less than that recorded on the Wilson frame (4.6°±3.1°, p<.001), and also significantly less than that recorded on the Jackson frame (p≤.001). Similar differences in segmental measurements were observed at L3–L4.
Conclusions
Prone positioning of patients in flexion on the operating table using the Andrews table or Wilson frame resulted in similar lumbar interspinous distance compared with the modified lateral decubitus position. Prone positioning on the Jackson frame resulted in statistically less interspinous distance than all other positions. Positioning on the Andrews table resulted in similar segmental angulation to the modified lateral decubitus position. Extrapolation from these data, obtained in healthy males younger than the typical age of patients treated with interspinous distraction devices, should clearly be done with caution. However, it seems reasonable to suggest that performing these procedures in the prone position using the Andrews table (greatest interspinous distance) is unlikely to result in the placement of significantly undersized implants, or significantly increase the force required to insert an implant.
Keywords: Interspinous process decompression, Lumbar spine, Lateral film, Spinous process distance, Vertebral disc angulation, Operative position
To access this article, please choose from the options below
FDA device/drug status: not applicable.
Author disclosures: JBT (stockholder, manager for medical affairs, Medtronic Spine LLC); WLP (stockholder, data and statistics director at Medtronic Spine LLC); TA (royalties, Medtronic Spine; stockholder at Simpirica Spine; consultant for Medtronic and Simpirica; member of board of directors and scientific advisory board of Simpirica Spine; research support from SPO #38201).